Status and Event Reporting System
VX4101A MultiPaq Instrument User Manual
4–11
STATus : OPERation : NTRansition
STATus:OPERation:NTRansition
STATus:OPERation:NTRansition?
Instrument
0
0–32767
N/A
Sets the Operational Negative Transition Filter for the currently selected
instrument. Setting a bit in this filter latches one to zero transitions of the
corresponding bit in the Operational Condition Register into the Operational
Event Register.
Queries the Operational Negative Transition Filter for the currently selected
instrument. The response is in NR1 format with a range from 0-32767.
Command Syntax
Query Syntax
Command Class
*RST Value
Limits
Related Commands
Command Description
Query Response