IEEE-488.2 Common Commands
VX4101A MultiPaq Instrument User Manual
3–303
This command is equivalent to a Group Execute Trigger command. Upon its
receipt, places any instrument which has selected BUS as its trigger source in the
Device Trigger Active State as defined by the IEEE 488.2 standard.
Command Response
INST:SEL DMM
CONF:VOLT:DC
TRIG:SOUR BUS
INIT
FETC:COUN? 0
*TRG
FETC:COUN? 1
FETC? +2.39340E-1
*TST?
N/A
*TST?
Global
N/A
N/A
TEST:ALL?
This query is used to perform a system wide self-test of the VX4101A infrastruc-
ture and all associated instruments. At the end of the self-test the overall pass/fail
status, as well as the pass/fail status for each instrument is returned as a response
to the query. If a test fails, a descriptive error message is placed in the error
queue and the error LED begins to blink. At the end of the self-test, the
VX4101A is returned to the reset condition. The reponse format is as follows:
Command Description
Examples
Command Syntax
Query Syntax
Command Class
*RST Value
Limits
Related Commands
Query Response