SCPI Commands for the Digital Input
VX4101A MultiPaq Instrument User Manual
3–119
CALibration Subsystem
The CALibration subsystem handles the calibration operations of the instrument.
CALibration:VALue <cal value>
N/A
N/A
Parameter Value
<cal value> 2.5|12
N/A
This command calibrates the Digital Input threshold. This is a 2-point calibra-
tion. The voltages used for calibration are 2.5 VDC and 12 VDC. You must use
external 2.5 V and 12 V sources to perform the calibration. You must hook up
the external source to Discrete Input 1 and set the Digital Output source to 1
(inactive).
NOTE. The Digital Output instrument uses the same pins as the Digital Input, so
the Digital Output should be reset prior to calibrating the Digital Input to ensure
that the outputs are inactive.
N/A
Command Response/Description
CAL:VAL 2.5 Sets the 2.5 V calibration factor using an external 2.5 V source.
Command Syntax
Query Syntax
*RST Value
Limits
Related Commands
Command Description
Query Response
Examples