Appendix G: Calibration
VX4101A MultiPaq Instrument User Manual
G–7
This part of the procedure calibrates the DMM for AC/DC Coupled Range
3 VDC or RMS.
11. Program the VX4101A as follows:
conf:acdc 3;init
12. Apply .300 VRMS ±5.3 mVRMS at 1 kHz
13. Program the VX4101A as follows:
cal:val .3 (or any other desired value)
14. Apply 2.90 VRMS ±35 mVRMS at 1 kHz
15. Program the VX4101A as follows:
cal:val 2.9 (or any other desired value)
This part of the procedure calibrates the DMM for AC/DC Coupled Range
30 VDC or RMS.
16. Program the VX4101A as follows:
conf:acdc 30;init
17. Apply 3.0 VRMS ±35 mVRMS at 1 kHz
18. Program the VX4101A as follows:
cal:val 3 (or any other desired value)
19. Apply 29.0 VRMS ±450 mVRMS at 1 kHz
20. Program the VX4101A as follows:
cal:val 29 (or any other desired value)
This part of the procedure calibrates the DMM for AC/DC Coupled Range
300 VDC or RMS.
21. Program the VX4101A as follows:
conf:acdc 300;init
22. Apply 30.0 VRMS ±450 mVRMS at 1 kHz
23. Program the VX4101A as follows:
cal:val 30 (or any other desired value)
24. Apply 290 VRMS ±4.5 mVRMS at 1 kHz
AC/DC Coupled Range
3 VDC or RMS
AC/DC Coupled Range
30 VDC or RMS
AC/DC Coupled Range
300 VDC or RMS