Tektronix VX4101A Drums User Manual


 
SCPI Commands for the Digital Multimeter
3–232
VX4101A MultiPaq Instrument User Manual
Parameter Value
<array_size> 1–4096
<expected_value> See tables 3-7 - 3-10
NOTE. The <array_size> limit is 2048 when the aperture time is one second.
CONFigure?
INPut
SENSe
READ?
INITiate
FETCh?
CALibrate:LFRequency
CALibrate:ZERO:AUTO
N/A
MEASure([:SCALar]|:ARRay)[:VOLTage][:DC]? [<Array Size>[,<Expected Value>]]]
Performs DC voltage measurements.
MEASure([:SCALar]|:ARRay)[:VOLTage]:AC? [<Array Size>[,<Expected Value>]]]
Performs AC coupled, AC TRMS voltage measurements. The AC bandwidth is
reset to slow mode.
MEASure([:SCALar]|:ARRay)[:VOLTage]:ACDC? [<Array Size>[,<Expected Value>]]]
Performs DC coupled, AC TRMS voltage measurements. The AC bandwidth is
reset to slow mode.
MEASure([:SCALar]|:ARRay):CURRent[:DC]? [<Array Size>[,<Expected Value>]]]
Performs DC current measurements.
MEASure([:SCALar]|:ARRay):RESistance? [<Array Size>[,<Expected Value>]]]
Performs two-wire resistance measurements.
MEASure([:SCALar]|:ARRay):FRESistance? [<Array Size>[,<Expected Value>]]]
Performs four-wire resistance measurements.
Limits
Related Commands
Command Description
Query Response