SCPI Commands for the Counter
3–42
VX4101A MultiPaq Instrument User Manual
Query Response
ARM(:SEQuence2 |:STOP)
[:LAYer]:SOURce:CATalog[:
ALL]?
Lists all trigger sources available for use with the
ARM:STOP:SOUR command. This command specifies the
stop arming source for the Counter. The possible responses
are as follows:
HOLD, IMMEDIATE, BUS, TTLTRG0, TTLTRG1, TTLTRG2,
TTLTRG3, TTLTRG4, TTLTRG5, TTLTRG6, TTLTRG7,
COMMAND0, COMMAND1, COMMAND2, COMMAND3,
COMMAND4, TIMER, SUREPATH, DMM, COUNTER,
CTR_EXTARM, DAC, DIGI, DIGO, INTERNAL, LEVEL
ARM(:SEQuence2 |:STOP)
[:LAYer]:SOURce:CATalog:
DELayable?
Lists all delayable trigger sources available for use with the
ARM:STOP:SOUR command. This command specifies the
stop arming source for the Counter. The possible responses
are as follows:
BUS, EXTERNAL, TTLTRG0, TTLTRG1, TTLTRG2, TTLTRG3,
TTLTRG4, TTLTRG5, TTLTRG6, TTLTRG7, COMMAND0,
COMMAND1, COMMAND2, COMMAND3, COMMAND4,
SUREPATH, DMM, COUNTER,
CTR_EXTARM, DAC, DIGI, DIGO
ARM (:SEQuence2 |:STOP)
[:LAYer]:SOURce:CATalog:
FIXed?
Lists all fixed trigger sources available for use with the
ARM:STOP:SOUR command. This command specifies the
stop arming source for the Counter. The possible responses
are as follows:
HOLD, IMMEDIATE, TIMER, INTERNAL, LEVEL
ARM ([:SEQuence1] |:STARt) [:LAYer]:DELay <delay in seconds>
Command Description
INST:SEL COUNTER Selects the Counter
CONF:VOLT:DC Sets the Counter for DC voltage measurements
ARM:SOUR TTLT0 Arms the TTLT0 source
ARM:DEL 1E–3 Sets time delay for 1E–3 seconds
INIT Begins the measurement
ARM ([:SEQuence1] |:STARt) [:LAYer]:ECOunt
Command Response
INST:SEL COUNTER Selects the Counter
CONF:VOLT:DC Sets the Counter for DC voltage measurements
ARM:SOUR TTLT0 Arms the TTLT0 source
ARM:ECO 100 Sets the event counter for 100 events
INIT Begins the measurement
Examples