Tektronix VX4101A Drums User Manual


 
SCPI Commands for the SurePatht Modules
3–290
VX4101A MultiPaq Instrument User Manual
TRIGger([:SEQuence]|:STARt)[:LAYer]:ECOunt <triggers to count>
Specifies the number of triggers to count prior to triggering. Upon receipt of
trigger N (where N is the number specified in the command), the instrument will
enter the triggered state. If the trigger source selected is fixed, this command will
have no effect on the instrument triggering. This command always zeros the
delay by time parameter, so specifying an event count of zero places the trigger
subsystem in pass-through mode. In this mode, the instrument triggers immedi-
ately upon receipt of a trigger.
TRIGger([:SEQuence]|:STARt)[:LAYer]:IMMediate
Causes a one time entry into the triggered state without receiving the specified
trigger. This command is often used to prepare for cases such as setting up a scan
list measurement. The example below shows an example of an automated
integrated trigger scan of four resistance values in autorange.
TRIGger([: SEQuence] |:STARt) [:LAYer]:SOURce <source>
Selects or queries the trigger source to be used when the instrument is initiated.
Command Response
TRIGger([:SEQuence]|:
STARt) [:LAYer]:DELay
<delay in seconds>
TRIGger([:SEQuence]|:
STARt)[:LAYer]:ECOunt
<triggers to count>
TRIGger([:SEQuence]|:
STARt) [:LAYer]:SOURce
<current source>
TRIGger([:SEQuence] |:
STARt)[:LAYer]:SOURce:
CATalog[:ALL]?
Lists all trigger sources available for use with the TRIG:SOUR
command.
HOLD, IMMEDIATE, BUS, TTLTRG0, TTLTRG1, TTLTRG2,
TTLTRG3, TTLTRG4, TTLTRG5, TTLTRG6, TTLTRG7,
COMMAND0, COMMAND1, COMMAND2, COMMAND3,
COMMAND4, TIMER, SUREPATH, DMM, COUNTER,
CTR_EXTARM, DAC, DIGI, DIGO
TRIGger([:SEQuence]|:
STARt)[:LAYer]:SOURce:
CATalog:DELayable?
Lists all delayable trigger sources available for use with the
TRIG:SOUR command.
BUS, TTLTRG0, TTLTRG1, TTLTRG2, TTLTRG3, TTLTRG4,
TTLTRG5, TTLTRG6, TTLTRG7, COMMAND0 COMMAND1,
COMMAND2, COMMAND3, COMMAND4, SUREPATH, DMM,
COUNTER, CTR_EXTARM, DAC, DIGI, DIGO
TRIGger([:SEQuence]|:
STARt)[:LAYer]:SOURce:
CATalog:FIXed?
HOLD, IMMEDIATE, TIMER
Query Response