Using the Digital Output
VX4101A MultiPaq Instrument User Manual
2–39
Outputing a Sequence
This procedure will show you how to output a sequence of x segments of y
points with the Digital Output. The internal clock is set to 1 kHz.
1. Select the Digital Output:
INST:SEL DIGO
2. Define the number of data points:
TRACe (1:Y),<data 1,...,data Y>
3. Set the output sample frequency to 1 kHz:
TRACe:SRATe 1000
4. Set the segment length to the number of data points Y:
TRACe:Points Y
5. Set the sequence number:
TRIG:MODE SEQuence, <number of segments>
6. Initiate the configuration:
INIT:IMM