Tektronix VX4101A Drums User Manual


 
SCPI Commands for the Counter
VX4101A MultiPaq Instrument User Manual
3–89
INSTrument Subsystem
The Instrument subsystem controls the instrument’s status. You can use it to end
a measurement or to make the instrument available for a measurement.
INSTrument
:ABORt
:RESet
N/A
Instrument
N/A
N/A
ABORt
INITiate
INITiate:CONTinuous
The following summarizes the INSTrument subsystem functionality:
INSTrument:ABORt
Places active instrument in the IDLE state, aborting any measurement or other
instrument activity in progress. The instrument configuration is unchanged and a
subsequent INIT command will cause the instrument to re-start the same type of
measurement. If the instrument is in Asynchronous mode, this command can be
sent while a query is in progress and the measurement will be aborted.
If the instrument is in Synchronous mode, this command will be queued while a
query is in progress. This is a ramification of the IEEE 488.2 Message Exchange
Protocol Enforcer (MEPE).
INSTrument:RESet
This command resets the currently selected instrument without affecting other
instruments. The instrument returns to its *RST state. The instrument remains
selected.
N/A
Command Syntax
Query Syntax
Command Class
*RST Value
Limits
Related Commands
Command Description