Tektronix VX4101A Drums User Manual


 
SCPI Commands for the Digital Input
VX4101A MultiPaq Instrument User Manual
3–143
Query Response
TRIGger([:SEQuence1]|:
STARt)[:LAYer]:SOURce:
CATalog:DELayable?
Lists all delayable trigger sources available for use with the
TRIG:SOUR command.
BUS, EXTERNAL, TTLTRG0, TTLTRG1, TTLTRG2, TTLTRG3,
TTLTRG4, TTLTRG5, TTLTRG6, TTLTRG7, COMMAND0,
COMMAND1, COMMAND2, COMMAND3, COMMAND4,
SUREPATH, DMM, COUNTER, CTR_EXTARM, DAC, DIGI,
DIGO
TRIGger([:SEQuence1]|:
STARt)[:LAYer]:SOURce:
CATalog:FIXed?
Lists all fixed trigger sources available for use with the
TRIG:SOUR command.
HOLD, IMMEDIATE, TIMER, HANDSHAKE
TRIGger:MASK <mask>
Command Response
TRIG:MATC #h00000000 Sets the value of the match pattern for all input bits. All bits are
set to 0.
TRIGGER:MASK #hffffffff Sets the match mask to “care” bits.
TRIG:MASK? #hffffffff
Queries the mask
TRIGger ([:SEQuence1] |:STARt)[:LAYer]:DELay
Command Response
INST:SEL DIGIN
Selects the Digital Input instrument
CONF:VOLT:DC Configures the Digital Input for a DC voltage reading
TRIG:SOUR TTLT0 Sets the trigger source to be TTLT0
TRIG:DEL 1E-3 Sets the trigger delay to 1 mSec
INIT Initiates the measurement
TRIGger ([:SEQuence1] |:STARt)[:LAYer]:ECOunt
Command Response
INS:SEL DIGIN Selects the Digital Input instrument
CONF:VOLT:DC Configures the Digital Input for a DC voltage reading
TRIG:SOUR TTLT0 Sets the trigger source to be TTLT0
TRIG:ECO 100 Sets the trigger count to 100
INIT Initiates the measurement
Examples