Tektronix VX4101A Drums User Manual


 
SCPI Commands for the Digital Input
VX4101A MultiPaq Instrument User Manual
3–139
N/A
*TST?
Initiates the Digital Input self test operation and returns one of two possible
responses:
“DIGI: Self Test Passed”
“DIGI: Self Test Failed”
NOTE. If the self-test fails, you can obtain further information with a SYST:ERR?
query.
Command/Query Response/Description
INSTRUMENT:SELECT DIGI Selects the Digital Input instrument.
TEST:ALL? ”Digital Input passes self-test.”
Runs the self-test.
TRIGger Subsystem
TRIGger:MASK <mask>
TRIGger:MATCh <pattern>
TRIGger([:SEQuence1]|:STARt)[:LAYer]:DELay <time delay>
TRIGger([:SEQuence1]|:STARt)[:LAYer]:ECOunt <event delay>
TRIGger([:SEQuence1]|:STARt)[:LAYer]:HANDshake:POLarity
:REQuest NORMal|INVerted <polarity>
:STRobe NORMal|INVerted <polarity>
TRIGger([:SEQuence1]|:STARt)[:LAYer]:IMMediate
TRIGger([:SEQuence1]|:STARt)[:LAYer]:MODE <mode>
TRIGger([:SEQuence1]|:STARt)[:LAYer]:SOURce <source>
TRIGger:MASK?
TRIGger:MATCh?
Limits
Related Commands
Query Response
Examples
Command Syntax
Query Syntax