SCPI Commands for the Digital Multimeter
VX4101A MultiPaq Instrument User Manual
3–227
Command Response
FETCh? #239+1.23456E+00,+2.34567E+00<LF>
FETCh? 1 +1.23456E+00<LF>
FETCh? 2,2 #226+2.34567E+00,+3.45678E+00<LF>
FETCh? 2,1,2 #226+1.23456E+00,+3.45678E+00<LF>
Fetch:count? 3<LF>
INITiate Subsystem
The INITiate commands initiate the DMM to begin making a measurement.
INITiate[:IMMediate]
INITiate:CONTinuous [<Boolean>]
INITiate:CONTinuous?
Parameter Value
<Boolean> 0 (OFF)
Parameter Value
<Boolean> ON |1| OFF |0
MEASure?
CONFigure
SENSe
FETCh?
TRIGger
ABORt
INITiate:CONTinuous [Boolean]
The INITiate:CONTinuous ON command internally triggers the DMM to operate
in a free-running mode, where it continuously acquires and updates a single
measurement. The FETCh? command can be issued any time after this command
to read the current value of the input. The INITiate:CONTinuous command is
independent of the trigger setup.
Examples
Command Syntax
Query Syntax
*RST Value
Limits
Related Commands
Command Description