Tektronix VX4101A Drums User Manual


 
SCPI Commands for the Counter
3–92
VX4101A MultiPaq Instrument User Manual
The aperture for totalize is 9E6 seconds. An ABORt command must be issued to
terminate the measurement.
NOTE. The Counter will perform an implied abort for a measurement query that
is received during a measurement. The measurement in progress will be
terminated and the measure command will be executed.
You can program the Counter to make and store up to 1000 measurements. The
SCALar commands program the Counter for one measurement. The ARRAy
commands program the Counter for 1 to 1000 measurements as specified by the
<array size> parameter.
N/A
MEASure[1|2|3]([:SCALar]|:ARRay)
:FREQuency? [<array size>[,<expected value>[,resolution>]]]
:FREQuency:RATio? [<array size>[,<second channel>[,<expected
value>[,resolution>]]]]
:PERiod? [<array size>[,<expected value>[,resolution>]]]
MEASure[1|2]([:SCALar]|:ARRay)
:NDUTycycle|PDUTycycle|DCYCle? [<array
size>[,<reference>[,<expected value>[,resolution>]]]]
:NWIDth|PWIDth? [<array size>[,<reference>[,<expected
value>[,resolution>]]]]
:PHASe? [<array size>[,<expected value>[,resolution>]]]
:RTIMe|FTIMe|RISE:TIME|FALL:TIME? [<array size>[,<low
reference>[,<high reference> [,<expected
value>[,resolution>]]]]]
:TINTerval? [<array size>[,<expected value>[,resolution>]]]
[:VOLTage]
:AC? [<array size>[,<expected value>[,resolution>]]]
:DC? [<array size>[,<expected value>[,resolution>]]]
:MINimum? [<array size>[,<expected value>[,resolution>]]]
:MAXimum? [<array size>[,<expected value>[,resolution>]]]
Command Syntax
Query Syntax