Tektronix VX4101A Drums User Manual


 
SCPI Commands for the Digital to Analog Converter
VX4101A MultiPaq Instrument User Manual
3–181
TRACE|DATA[:DATA]? <Channel Select>|<Numeric List of Indices>
TRACE|DATA[:DATA]? <Channel Select>
TRACe:INDice?
TRACe:POINts?
TRACe:RFRequency?
TRACe:RPERiod?
TRACe:SRATe?
FORMat
Parameter Value
<trace voltages> 0.0
TRACe:INDice? 0,0
NOTE: At reset, the DAC clears its memory and returns a segment index of 0. The DAC will not
perform an output operation and remains at the 0,0 indice until you send a TRACe|DATA
command and the instrument is initiated.
<Trace Buffer Points> 1
<repeat frequency> OFF
<repeat period> OFF
<sample_rate> 15000 Hz
Parameter Limit
<Channel>,<Channel Select> 1-8
<Index> 1 v index v 1024
<Trace Voltage> –14.0 v voltage value v + 14.0
TRACe:INDice? 1 v sement v 4096
1 v sample index v 1024
<Trace Buffer Points> 1 v Trace Buffer Points v 1024
and
(Trace Buffer Points/Sample Rate) v 16 seconds
<repeat frequency> .0625 Hz v Waveform Frequency v 15000 Hz or OFF
<repeat period> 66.67E–6 seconds v Waveform period v 16 seconds, or
OFF
<sample_rate> 3.662 Hz v Sample Rate v 15000 Hz and (Trace Buffer
Points/Sample Rate) v 16 seconds
Query Syntax
Related Commands
*RST Value
Limits