The STATus:OPERation:CONDition query returns the current operational status
of the DMM board. The bit definitions of the value are (bit 0 = the least
significant bit):
Bit Definition Function
0 Calibrating Set when any CALibration operation is running. Cleared when
the CALibration operation is complete
1 Settling Set when the instrument changes its function or range. Cleared
when the all circuitry has settled
2 Ranging Set when the instrument is autoranging. Cleared when the
input range has been found
3 Sweeping Not used
4 Measuring Set when the INITiate command is executed. Cleared when the
command is complete or aborted
5 Triggering Set when the instrument is waiting for a trigger signal. Cleared
when the trigger is received
6 Arming Not used
7 Correcting Set when the instrument is performing an autozero operation.
Cleared when the autozero operation is complete
8 Testing (User 1) Set when the instrument is performing a self-test. Cleared
when the self-test is complete
9 Aborting (User 2) Set when the instrument is in the process of aborting an
operation. Cleared when the abort is complete
10 User 3 Not used
11 User 4 Not used
12 User 5 Reserved
13 Instrument Summary Not used
14 Program Running Not used
15 Reserved Always 0
Command Response
status:operation:condition? 16 Measurement in progress (0010 hex).
stat:oper:cond? 3072 Measurement complete because an ABORt was received
(0C00 hex).
Query Response
Examples