Tektronix VX4101A Drums User Manual


 
SCPI Commands for the Digital to Analog Converter
VX4101A MultiPaq Instrument User Manual
3–173
TRACe
The command specifies how data is returned from a waveform (TRACe
subsystem) voltage value query. Data can be returned in ASCII or binary format.
The query returns the currently selected response data format.
Command Response/Description
FORMat BINary N/A
Specifies the use of binary when returning trace voltage values.
The binary format of the return trace data is based upon the
IEEE 488.2 Indefinite Arbitrary Block Program Data type.
The output amplitude = (x–2048) * 8 mV
where x=256*MSByte+LSByte.
INITiate Subsystem
These commands cause the current TRIGger command sequence to begin.
INITiate
:CONTinuous [0|OFF|1|ON]
[:IMMediate]
INITiate:CONTinuous?
Instrument
N/A
0, 1, ON, or OFF
ABORt
INSTrument:ABORt
INITiate:CONTinuous [0|OFF|1|ON]
This command initiates the current programmed trigger sequence. After the
instrument has completed the current trigger sequence, it enters the initiated state
Related Commands
Command Description
Query Response
Examples
Command Syntax
Query Syntax
Command Class
*RST Value
Limits
Related Commands
Description