SCPI Commands for the Digital Multimeter
VX4101A MultiPaq Instrument User Manual
3–253
TRIGger ([:SEQuence1] |:STARt)[:LAYer]:DELay
Command Response
INST:SEL DMM Selects the DMM
CONF:VOLT:DC Configures the DMM for DC measurements
TRIG:SOUR TTLT0 Selects TTLT0 as the trigger source
TRIG:DEL 1E-3 Sets the trigger delay
INIT Initiates the measurement
TRIGger ([:SEQuence1] |:STARt)[:LAYer]:ECOunt
Command Response
INS:SEL DMM Selects the DMM
CONF:VOLT:DC Configures the DMM for DC measurements
TRIG:SOUR TTLT0 Selects TTLT0 as the trigger source
TRIG:ECO 100 Sets the event count to 100
INIT Initiates the measurement
TRIGger ([:SEQuence1] |:STARt)[:LAYer]:IMMediate
Command Response
TRIG:SOUR:CAT? IMMEDIATE, BUS, TTLTRG0, TTLTRG1, TTLTRG2,
TTLTRG3, TTLTRG4, TTLTRG5, TTLTRG6, TTLTRG7,
COMMAND0, COMMAND1, COMMAND2, COMMAND3,
COMMAND4, TIMER, SUREPATH, DMM, COUNTER,
CTR_EXTARM, DAC, DIGI, DIGO, HANDSHAKE
INST:SEL DMM Selects the DMM
CONF:VOLT:DC Configures the instrument for DC measurements
TRIG:SOUR TTLT0 Selects TTLT0 as the trigger source
INIT Initiates the measurement
FETC:COUN? 0
Returns the number of measurements completed
TRIG:IMM Sets triggering to immediate
FETC:COUN? 1
Returns the number of measurements completed. This could
vary depending on the actual measurements.
F? –6.03720E+00
Examples