Tektronix VX4101A Drums User Manual


 
SCPI Commands for the Digital Multimeter
3–230
VX4101A MultiPaq Instrument User Manual
N/A
The related commands of INSTrument subsystem are as follows:
ABORt
INITiate
INITiate:CONTinuous
INSTRument:ABORt
This command places the active instrument in the IDLE state, aborting any
measurement or other instrument activity in progress. The instrument configura-
tion is unchanged and a subsequent INIT command will cause the instrument to
restart the same type of measurement.
NOTE. If the instrument is in asynchronous mode, this command can be sent
while a query is in progress and the measurements will be aborted. If however,
the instrument is in synchronous mode, this command will be queued while a
query is in progress.
INSTrument:RESet
This command resets the currently selected instrument without effecting other
instruments. The instrument returns to its *RST state. The instrument remains
selected.
N/A
INSTRument:ABORt
Command Response
INST:SEL DMM
CONF:ARR:VOLT:DC 512
INIT
FETC:COUN? 127
INST:ABOR
FETC:COUN? 153
NOTE: After an ABORt command, the instrument stops taking measurements.
Limits
Related Commands
Command Description
Query Response
Examples