Tektronix VX4101A Drums User Manual


 
SCPI Commands for the Counter
3–60
VX4101A MultiPaq Instrument User Manual
correction factors for the Counter attenuation and gain settings. This command
requires an input calibration source set to the <input voltage> values below
(
±0.1%).
Table 3–3: Input Calibration Source Settings
<input voltage>
+50.0
+20
+10
+5.0
+2.0
+1.0
+0.5
+0.2
+0.1
+0.05
–50.0
–20
–10
–5.0
–2.0
–1.0
–0.5
–0.2
–0.1
–0.05
[<channel>] specifies the channel being calibrated, and can be specified as either
1 or 2. If [<channel>] is not specified, channel one is assumed.
CALibrate[<channel>]:ZERO
The CALibrate[<channel>]:ZERO command performs offset corrections for each
of the gain and attenuation settings of the Counter. This command requires an
input calibration source set to 0.0 V ±0.001 V.
[<channel>] specifies the channel being calibrated, and can be specified as either
1 or 2. If [<channel>] is not specified, channel one is assumed.