Tektronix VX4101A Drums User Manual


 
SCPI Commands for the Counter
3–88
VX4101A MultiPaq Instrument User Manual
INPut[1|2]:OFFSet[:ABSolute]
The channel one input signal is a 0.5 V
p-p
sine wave with a +.25 VDC compo-
nent and the user wishes to remove the DC component with the INPut:OFFset
command.
Command Response
INSTrument:SELect Counter Selects the Counter
INPut:SETup:AUTo off
SENSe:FUNCtion
“FREQuency”
INPut:COUPling DC
INPut:OFFSet .25
INPut[1|2]:COMParator[1|2]:LEvel[:ABSolute]
In this example, the channel one input signal varies between +0.5 and –0.5 V and
you will cause the comparator to trigger at 0.25 V.
Command Response
INSTrument:SELect Counter
INPut:SETup:AUTo off
SENSe:FUNCtion
“FREQuency”
INPut:COUPling DC
INPut:COMParator:LEVel .25
INPut[1|2]:OFFSet:RELative
The channel one input signal is a 0.5 V
p-p
sine wave with a +0.25 VDC
component. The user wishes to remove the DC component with the INPut:OFF-
set:RELative command.
Command Response
INSTrument:SELect Counter
INPut:SETup:AUTo off
SENSe:FUNCtion
“FREQuency”
INPut:COUPling DC
INPut:OFFSet:RELative .25
Examples