Appendix G: Calibration
VX4101A MultiPaq Instrument User Manual
G–9
10. Program the VX4101A as follows:
cal:val 10 (or any other desired value)
This part of the procedure calibrates the DMM for 300 W resistance.
11. Program the VX4101A as follows:
conf:fres 300;init
12. Apply 0.0 W ±1.0 mW
13. Program the VX4101A as follows:
cal:val 0
14. Apply 100.0 W ±3 mW
15. Program the VX4101A as follows:
cal:val 10 (or any other desired value)
This part of the procedure calibrates the DMM for 300 kW resistance.
16. Program the VX4101A as follows:
⊂onf:fres 3000;init
17. Apply 0.0 W ±4.0 mW
18. Program the VX4101A as follows:
cal:val 0
19. Apply 1000.0 W ±16 mW
20. Program the VX4101A as follows:
cal:val 100 (or any other desired value)
This part of the procedure calibrates the DMM for 30 kW resistance.
21. Program the VX4101A as follows:
conf:fres 30000;init
22. Apply 0.0 W ±40 mW
23. Program the VX4101A as follows:
cal:val 0
Resistance 300 W
Resistance 3 kW
Resistance 30 kW