Tektronix VX4101A Drums User Manual


 
SCPI Commands for the Digital Multimeter
3–208
VX4101A MultiPaq Instrument User Manual
Threshold
2
Threshold
2
u Threshold
1
Threshold
1
Threshold
2
Threshold
1
Threshold
2
t Threshold
1
Threshold
2
u Threshold
1
CALCulate:LIMit:LOWer[DATA] <threshold>
This command sets a lower threshold limit. This command is different from the
envelope command in that it checks only a lower threshold
CALCulate:LIMit:UPPer[DATA] <threshold>
This command sets an upper threshold limit. This command is different from the
envelope command in that it checks only an upper threshold.
CALCulate:TRANsform:HISTogram:COUNt <numeric value>
This command determines the number of data points to include in a histogam
calculation. The query version returns the programmed count.
CALCulate:TRANsform:HISTogram:POINts <numeric value>
This command sets the number of intervals in a histogram calculation. If a data
value is exactly the same value as a limit dividing two bins, the data value will
be counted in the next bin, unless the value matches the last interval limit. The
query version returns the number of intervals programmed.
Bin # Limits
1 5 –10
2 10 –15
3 15 –20
4 20 –25
5 25 –30
6 30 –35