SCPI Commands for the Digital Multimeter
VX4101A MultiPaq Instrument User Manual
3–237
Command/Query Response/Description
meas:resistance?
CONF?
+12.3456<LF>
:SCAL:RES 30,0.0001
Perform a 2-wire, autorange resistance measurement. The
default resolution is used to set the aperture to 200 ms. Issue
the following commands:
Meas:arr:curr? 3,0.15,
maximum
conf?
+0.109876,+0.109678,+0.109444<LF>
:ARR:CURR:DC 3,0.15,1.58114e-07
Perform three current measurements on the 150 mA range.
Specifying the maximum resolution sets the aperture value to 2
seconds. Issue the following commands:
TEST Subsystem
The TEST subsystem handles the self test operations of the instrument. The
Digital Output self test tests the Digital Output memory and the Read/Write
hardware control registers. The query returns pass/ fail information. In a failed
situation, additional failure information can be obtained with the SYStem:ER-
Ror? query.
N/A
TEST:ALL?
N/A
N/A
*TST?
Initiates the Digital Output self test operation and returns one of two possible
responses:
“DIGO Self-Test Passed”
“DIGO Self-Test Failed”
Command Syntax
Query Syntax
*RST Value
Limits
Related Commands
Command Description