Tektronix VX4101A Drums User Manual


 
Appendix G: Calibration
VX4101A MultiPaq Instrument User Manual
G–3
conf:dc .3;init
8. Apply 0.0 VDC ±1 mVDC
9. Program the VX4101A as follows:
cal:val 0
10. Apply 0.290 VDC ±5.3 mVDC
11. Program the VX4101A as follows:
cal:val .29 (or any other desired value)
This part of the procedure calibrates the DMM for 3 VDC with 10 MW
impedance.
12. Program the VX4101A as follows:
conf:dc 3;init
13. Apply 0.0 VDC ±4 mVDC
14. Program the VX4101A as follows:
cal:val 0
15. Apply 2.90 VDC ±35 mVDC
16. Program the VX4101A as follows:
cal:val 2.9 (or any other desired value)
This part of the procedure calibrates the DMM for 30 VDC with 10 MW
impedance.
17. Program the VX4101A as follows:
conf:dc 30;init
18. Apply 0.0 VDC ±40 mVDC
19. Program the VX4101A as follows:
cal:val 0
20. Apply 29.0 VDC ±450 mVDC
21. Program the VX4101A as follows:
cal:val 29 (or any other desired value)
DC Range 3 VDC with
10 MW Impedance
DC Range 30 VDC with
10 MW Impedance