Tektronix VX4101A Drums User Manual


 
SCPI Commands for the Digital Multimeter
3–228
VX4101A MultiPaq Instrument User Manual
The INITiate:CONTinuous OFF command aborts any measurements currently in
progress.
INITiate[:IMMediate]
Instrument initiates its current trigger sequence. After the instrument has
completed the current trigger sequence, it enters the idle state. See instrument
documentation for details on instrument state after an initiate.
NOTE. If data sampling is in progress, issuing an INITiate command aborts the
acquisition and starts a new acquisition.
INITiate:CONTinuous?
Returns a zero or the number one indicating whether INITiate:CONTinuous is
off or on.
Command Response
INITiate:IMMediate Arms the triggers for an acquisition
init:continuous Put the DMM into free-running acquisitions
INPut Subsystem
The INPut command defines the input impedance for the lower three DC ranges
of the DMM.
INPut:IMPedance <impedance>
INPut:IMPedance?
Parameter Value
<impedance> 10e6 W
Parameter Value
<impedance> 10e6 W, 10e9 W
Query Response
Examples
Command Syntax
Query Syntax
*RST Value
Limits