SCPI Commands for the Counter
3–56
VX4101A MultiPaq Instrument User Manual
CALCulate:MINimum
Command Response
inst:sel ctr Selects the Counter
CALC:MINimum? Finds the minimum value
CALCulate:MAXimum
Command Response
inst:sel ctr Selects the Counter
CALC:MAXimum? Finds the maximum value
CALCulate:SDEViation
Command Response
inst:sel ctr Selects the Counter
CALC:SDEViation? Calculates and returns the standard deviation on a set of data
CALCulate:TRANsform:HISTogram
Command Response
inst:sel ctr Selects the Counter
CALC:TRAN:HIST:COUNt 5 Sets the number of data to use in a histogram calculation to 5
CALC:TRAN:HIST:POINts 3 Sets the number of intervals in a histogram calculation to 3
CALC:TRAN:HIST:RANGe
10,00
Sets the range of a histogram calculation
CALC:TRAN:HIST? Returns the histogram distribution
CALCulate:TRANsform:HISTogram:ABOVe?
Command Response
inst:sel ctr Selects the Counter
CALC:TRAN:HIST:RANGE:
AUTO OFF
Turns off autoranging
CALC:TRAN:HIST:RANGE:
10,50
Sets the range of the histogram calculation for 10-50
CALC:TRAN:HIST:ABOV? Returns the number of data points above 50