SCPI Commands for the Counter
VX4101A MultiPaq Instrument User Manual
3–49
range of the envelope is determined by the <threshold1> and the <threshold2>
values. For example:
H If the <threshold2> is greater than the <threshold1>, the range of data
values searched for is above the <threshold1> value or below the <thresh-
old1> value.
H If the <threshold2> is less than <threshold1>, the range of data values
searched for is between the <threshold1> and <threshold2>.
Threshold
2
Threshold
1
t Threshold
2
Threshold
1
Threshold
2
Threshold
1
Threshold
1
u Threshold
2
Threshold
1
t Threshold
2
CALCulate:LIMit:LOWer[:DATA] <threshold>
This command searches for all the input data values below a certain <threshold>.
CALCulate:LIMit:UPPer[:DATA] <threshold>
This command searches for all the input data values above a certain threshold.
CALCulate:TRANsform:HISTogram:COUNt <threshold>
This command determines the number of data points to include in a histogram
calculation.
CALCulate:TRANsform:HISTogram:POINts <threshold>
This command sets the number of intervals in a histogram calculation. If a data
value is exactly the same value as a limit dividing two bins, the data value will
be counted in the next bin, unless the value matches the last interval limit.
Bin # Limits
1 5 –10
2 10 –15
3 15 –20
4 20 –25